GE LOGIQ E9 Service Manual page 281

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

GE
D
5535208-100, R
IRECTION
7-5-6-3
Keyboard Tab
Test the keyboard functionality by typing on the keyboard. Verify that the corresponding key is
highlighted when each key is pressed. See:
NOTE:
The keyboard test does not give an indication when the Print Screen button is pressed.
. 2
EV
Figure 7-35 Keyboard Tab - R4 and earlier
Figure 7-36 Keyboard Tab - R5 and later
Chapter 7 Diagnostics/Troubleshooting
Figure 7-35
or
Figure 7-36.
LOGIQ E9 S
M
ERVICE
ANUAL
7 - 33

Advertisement

Table of Contents
loading

Table of Contents