GE
D
5535208-100, R
IRECTION
7-5-10-8
Network Adapter Diagnostic Changes for R4
Network Adapter Test is split into 2 folders:
Network Adapter - see:
Network Adapter Reliability - see:
Each folder contains separate tests for Wired and Wireless Adapters.
7-5-10-9
Memory
See:
7-5-3 "Diagnostics Window Overview" on page 7-22
Runs a test on essential functions of the Memory. The test results are displayed in the Status portion of
the window.
1.) Enter the desired Loop Count (numeric value of 4 digits or less).
2.) Click the Execute button.
3.) Replace the BEP if any tests fail.
. 2
EV
Table 7-124 on page 7-157
Chapter 7 Diagnostics/Troubleshooting
and
Table 7-125 on page 7-158
Table 7-126 on page 7-159
for a description of this window's features.
LOGIQ E9 S
ERVICE
and
Table 7-127 on page 7-160
M
ANUAL