D
5535208-100, R
IRECTION
8-6-15
Probe Cable Hooks replacement
8-6-15-1
Manpower
One person, 15 minutes.
8-6-15-2
Tools
For tools needed, refer to:
8-6-15-3
Probe Cable Hook removal
1.) Unscrew the fastening screw.
The screw may be locked with Lock-Tite, so you may need to apply a little extra force.
2.) Remove the Cable Hook.
8-6-15-4
Probe Cable Hook installation
1.) Position the Probe Cable Hook (Figure 8-55).
2.) Fix it in place by using an M5x20 screw with an M6 washer. Lock-Tite has to be used.
Torque: 50 Ncm (0.37 lbf-ft or 4.4 lbf-in) (fingertight).