7-5-6-9
Ending the Program
1.) When you have finished testing the Op Panel, close the GE Test App.
2.) Shutdown the system and reboot.
7-5-7
Diagnostics - Common Diagnostics
7-5-7-1
Disruptive Mode
The customer enables this feature by choosing "Disruptive Mode" and confirming "Yes", before a GE
Service FE can access the customer's ultrasound scanner remotely. "Disruptive Mode" can be
requested remotely by the service technician or OLC, or it can be selected by the customer directly on
the scanner or workstation.
Section 7-6 "Troubleshooting using InSite ExC" on page 7-161