Covers under XY / Frogleg motors removal
1.) At the rear of the system, release the console's frogleg mechanism by inserting a screwdriver into
the release point and pressing until release. Pull the console out to its extended position to gain
access to the screws in the next step.
2.) Under the XY / Frogleg mechanism, identify the four separate covers. There are two types of
covers, one type has two screws (1) and the other type has one screw (2).
3.) Remove the screws from the cover(s) to replace.
Figure 8-36 XY / Frogleg mechanism covers, right side frog leg, from underneath
. 2
EV
Figure 8-35 XY / Frogleg mechanism release