Keyboard (cont'd)
Footswitch Tests
Under Service Diagnostics / BEP Interactive/ Keyboard
INPUT > Press left, middle or right footswitch keys.
OUTPUT > Keys will be registered as follows:
-
-
-
7-5-12-2
Mouse (Trackball)
See:
7-5-3 "Diagnostics Window Overview" on page 7-22
Runs a test on essential functions of the Trackball. The test results are displayed in the Status portion
of the window.
1.) Enter the desired Loop Count (numeric value of 4 digits or less).
2.) Click the Execute button.
3.) Verify the signal cable is connected securely.
4.) Clean the Trackball. If the problem continues:
5.) Replace the Trackball.
. 2
EV
Left Footswitch = Left Windows + F10
Middle Footswitch = Left Windows + F11
Right Footswitch = Left Windows + F12
Chapter 7 Diagnostics/Troubleshooting
LOGIQ E9 S
for a description of this window's features.
M
ERVICE
ANUAL