10-4-5
Peripheral/option checks
If any peripherals or options are not part of the system configuration, the check can be omitted
Table 10-6
Step
1
2
3
4
5
6
10-4-6
Mains cable inspection
Table 10-7
Step
1
2
3
4
10-4-7
Optional diagnostic checks
To complete the Ultrasound System checks, view the error logs and run desired diagnostics.
10-4-8
View the logs
1.) Review the Ultrasound system error log for any problems.
2.) Check the temperature log to see if there are any trends that could cause problems in the future.
. 2
EV
GE approved peripheral/hardware option functional checks
Item
Media
Verify media drive(s) read/write properlty. Clean if necessary.
B/W Printer
Verify hardcopy output of the B/W video page printer. Clean heads and covers if necessary.
Color Printer
Verify hardcopy output of the Color video page printer. Clean heads and covers if necessary.
DICOM
Verify that DICOM is functioning properly. Send an image to a DICOM device.
ECG
Verify basic operation with customer
Footswitch
Verify that the footswitch is functioning as programed. Clean as necessary.
Mains Cable Inspection, As Appropriate
Item
Unplug cord
Disconnect the mains cable from the wall and Ultrasound system.
Inspect
Inspect it and its connectors for damage of any kinds.
Verify that the LINE, NEUTRAL and GROUND wires are properly attached to the terminals, and that
Verifyr
no strands may cause a short circuit.
Verify
Inlet connector retainer is functional.