Smart Power Supply Error Code 9 Flow Chart 1R "Belt Start Reject Error (9) - GE T2100-ST1 Service Manual

Hide thumbs Also See for T2100-ST1:
Table of Contents

Advertisement

Smart Power Supply Error Code 9 Flow Chart 1R "Belt Start
Reject Error (9)"
1R
Is the Red LED D13
flashing 9 times with 4
second pause on the
Smart Power Supply
PCBLF0443-1?
Yes
Turn the treadmill power switch
OFF (O).
No
Is the Red LED D13 on
the Smart Power
Supply PCBLF0443-1
on continuously?
Go To
Page 1I
Send a Start Belt Command or Press
Start if equipped with controller
increase speed to 10mph.
2097937-002 Rev G
11 March 2019
No
Inspect interface cable between
the Treadmill & Stress System
cables for damage.
Turn the treadmill power switch ON (-).
T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V
Yes
Go To
Page 1I
Is the Red LED D13
flashing on the Smart
Power Supply
PCBLF0443-1?
With the treadmill
running at 10.0 mph
for 1 minute is the
Red LED D13 on?
No
Go To
Page 1I
Troubleshooting
No
Treadmill should
be operational.
Is the RS232 or
USB Interface
cable damaged?
No
Replace
Interface
cable.
Yes
Return Treadmill
to service.
97