Table 3: Guidance And Manufacturer's Declaration – Immunity Me Equipment And Me Systems That Is Not Life-Supporting - GE T2100-ST1 Service Manual

Hide thumbs Also See for T2100-ST1:
Table of Contents

Advertisement

Introduction
Table 3: Guidance and Manufacturer's Declaration –
Immunity ME Equipment and ME Systems that is NOT
Life-supporting
Immunity
Test
Conducted
RF
IEC 61000-4-
6
Radiated RF
IEC 61000-4-
3
16
The T2100-ST Series is intended for use in the
electromagnetic environment specified below. The
customer or user of the T2100-ST Series should ensure
that it is used in such an environment.
IEC 60601
Test Level
3 Vrms
150 kHz to
80 MHz
3 V/m
80 MHz to
2.7 GHz
T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V
Compliance
Electromagnetic
Level
Environment - Guidance
3V, 6V at ISM
Portable and mobile
+ (Amateur
communications
Frequencies
equipment should be
separated from the
T2100-ST Series by no less
than the distances
3 V/m at 80 -
calculated/listed below:
2,700MHz,
AM
D=(3.5/V1) (Sqrt P)
Modulation
150kHz to 80MHz
9-28V/m, 385
- 6,000MHz,
FM or Digital
D=(3.5/E1) (Sqrt P)
Modulation
80 to 800 MHz
D=(7/E1 )(Sqrt P)
800 MHz to 2.5 GHz
where P is the max power
in watts and D is the
recommended separation
distance in meters.
Field strengths from fixed
transmitters, as
determined by an
electromagnetic site
survey, should be less
than the compliance
levels (V1 and E1).
Interference may occur in
the vicinity of equipment
containing a transmitter.
2097937-002 Rev G
11 March 2019

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

T2100-st2

Table of Contents