Calibration; Installing The T2100-St Series Calibration Software - GE T2100-ST1 Service Manual

Hide thumbs Also See for T2100-ST1:
Table of Contents

Advertisement

Calibration

9
Calibration is necessary to ensure your T2100-ST series treadmill
is functioning properly. This chapter describes how to:
• install the T2100-ST series calibration software,
• install the serial drivers,
• connect the calibration software to the T2100-ST series
• calibrate the treadmill's speed,
• calibrate the treadmill's elevation, and
• understand the calibration software's tabs.

Installing the T2100-ST Series Calibration Software

This section explains how to install the T2100-ST Series
Calibration Software on the field service engineer's PC in
preparation for calibrating the T2100-ST Series treadmill.
Compatible PC operating systems are Windows 7, 8, 8.1, and 10
(both 32 bit and 64 bit).
NOTE:
NOTE:
1.
2097937-002 Rev G
11 March 2019
treadmill,
This section can be skipped if the T2100-ST Series
Calibration Software is already installed on the PC being
used. If already installed, proceed to "Installing the Serial
Drivers" on page 163.
Do not plug any hardware including treadmill cable into the
PC unless directed to do so.
Download the T2100-ST Series Calibration software using
the following procedure:
a.
Go to the GE Service Support Central site;
T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V
9
161

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

T2100-st2

Table of Contents