Setup and Connection to Host

Currently, four GE host stress systems can connect to the T2100-
ST series treadmill and are used as controlling devices: CASE
CardioSoft/CS, MAC 5500, and MAC 2000.

CASE connection and setup:

1.
2.
2097937-002 Rev G
11 March 2019
Stage
Incline
13
0%
14
0%
NOTE:
Successful completion of the preceding testing
procedure ensures that the treadmill is fully
functional and responsive to command signals.
NOTE:
Unsuccessful completion of the preceding testing
procedure indicates a problem with the setup. Call GE
Medical Systems Information Technologies to trouble
shoot failure of test plug procedure.
Turn the power "OFF" at the treadmill.
Remove the test connector.
Reconnect the RS232 or USB interface cable from the host
computer.
You are ready to begin the set-up procedure prescribed by
your medical test equipment supplier.
Connect one end of the RS232 serial cable (GE part number
2097829-109) to the RS232 COM port located on the lower
front panel of the T2100-ST series treadmill and run it
through the communication cable clamp, as shown in the
following photograph.
Connect the other end of the RS232 serial cable (GE part
number 2097829-109) to a COM port on the CASE unit (it is
suggested to use COM1 or COM2).
T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V
Assembly and Setup