NOTE:
1.
2.
3.
4.
5.
6.
2097937-002 Rev G
11 March 2019
¾" Socket or Box Wrench
¾" Wrench
Tape measure
The drive belt may stretch and loosen with regular use. This
looseness may result in a flapping noise under the hood.
If there are indications that the drive belt has stretched and
become loose, contact your in-house biomedical
technician, or contact GE Medical Systems Information
Technologies, Inc., and open a customer service call.
Remove the (5) #10-32 screw located on the bottom hood
with
/
" Allen wrench.
1
8
With the ¾" socket and wrench loosen the TENSION
ADJUSTMENT back nut.
Press down on drive belt between motor and front roller
with approximately 5lbs of force to achieve ¼" to 3/8".
If adjustment is need using a ¾" wrench turn the TENSION
BOLT clockwise to tighten. To loosen, turn the wrench
counter clockwise.
Make small adjustment until the drive belt deflects
approximately ¼" to 3/8", tighten the TENSION
ADJUSTMENT back nut.
NOTE:
Failure to lock the TENSION ADJUSTMENT back nut
will allow the drive belt tension to become loose.
When the treadmill is properly adjusted, reinstall the hood
with (5) #8-32 screws.
1
2
Item
Description
1
¼ to 3/8 Deflection
2
5 lbs down force
T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V
Preventive Maintenance
3
61