H.
Emergency Stop Switch: Two 0.187" quick-connect blade
terminals are provided for an emergency stop switch input.

Physical Requirements and Restrictions

A.
The printed circuit board should be a rigid 1/16" thick, FR4,
SMOBC circuit board with 2-ounce copper.
B.
Galvanic isolation between the AC power circuits and the
low voltage electronic circuits should be guaranteed.
C.
Outer Dimensions: 6" T x 9" W x 2.5" D (max).
D.
Mounting position: The board shall be capable of being
mounted with the PCB either parallel or perpendicular to
the bottom pan of the treadmill.
E.
Environment: 0 - 70 °C ambient temperature.
F.
High level vibrations from adjacent equipment may be
present.
G.
After installation into the treadmill, expect the board to
experience moderate shock impulses during shipment.
2097937-002 Rev G
11 March 2019
T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V
Theory of Operation
73