Troubleshooting

Emergency Stop Flow Chart 1E

1E
Is the Red D13
LED lit on Smart
Power Supply
PCBLF0443-1?
No
Verify treadmill power switch is
in the off (0) position.
Verify the Emergency Stop
Button is released.
Turn treadmill power switch in
the on (1) position.
82
LED lit on Smart
Yes
Verify wire connectors on the Emergency
Go To
Chart 1F
Stop and quick disconnect on connector
J4 & J5 are a secure connection on Smart
Power Supply PCBLF0443-1.
Replace
Emergency
Stop Button.
T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V
Is the Red D13
Power Supply
PCBLF0443-1?
Verify the continuity of the Emergency
Yes
Stop Harness with an OHM meter to
Go To
ensure the wire is not damaged.
Chart 1F
Is the Red D13
LED lit on Smart
Power Supply
PCBLF0443-1?
Yes
No
No
Verify treadmill power switch is
in the off (0) position.
No
Wires
Damaged?
Stop Harness.
Turn treadmill power switch in
the on (1) position.
Red D13 LED Light should be
lit on Smart Power Supply
PCBLF0443-1 Reference 1E.
2097937-002 Rev G
11 March 2019
Yes
Replace
Emergency