Component/FRU Removal and Replacement

Replacing the Running Belt

Tools Required
o
o
o
o
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
134
¼" Allen wrench or socket
3/16" Allen wrench or socket
½" Wrench or socket with extension
Jack stand or block capable of supporting the weight of the rear of
the treadmill
Approximate time to complete: 1 hr 30 min
Elevate the treadmill to 12%.
Turn the treadmill main power switch to the OFF position.
Unplug the treadmill plug from the wall receptacle.
Allow the treadmill to discharge for 1 minute.
Remove the (5) hood-securing bolts.
Remove the hood and set it aside.
Lift the rear of the treadmill and place jack stands or a
block to support it.
With the 3/16" Allen wrench or socket, remove the bolt
securing the left and right end caps from the rear of the
treadmill.
Remove the end caps and set them aside.
With the ¼" Allen wrench or socket, completely remove the
left and right tension bolts from the rear roller.
When removing the tension bolts, maintain equal tension
on the bolts to avoid stripping the threads on the tension
bolt or roller.
Push the rear roller forward as far as it will go.
Lift the edge of the running belt up and use the 3/16" Allen
wrench or socket to remove the (4) flat head Allen screws
that secure the running deck cross braces.
There are (4) screws total, (2) on the left side and (2) on the
right side.
Push the deck to one side of the treadmill and, from the
other side, lift the edge of the deck up and out over the rail
and set it aside.
With the ¼" Allen wrench or socket, remove the bolt
securing the front roller to the treadmill frame.
Push the front roller from the right side of the treadmill to
the left side, lift the roller up, slip the timing belt off the
T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V
2097937-002 Rev G
11 March 2019