Communication Usb Flow Chart 1H - GE T2100-ST1 Service Manual

Hide thumbs Also See for T2100-ST1:
Table of Contents

Advertisement

Communication USB Flow Chart 1H

1H
When connecting Stress
System or Full Vision
Treadmill PS New Cal
Software, you received
missing driver error code.
No
When connecting Full
Vision Treadmill PS
New Cal Software are
you Recd and Sent
Yes
information?
Go To
Chart 1I
No
Verify treadmill power switch
is in the off (0) position.
Verify the Emergency Stop
Button is released or Pull
Tether missing.
Turn treadmill power switch
in the on (1) position.
Press connect on the
Full Vision Treadmill
PS New Cal Software
are you Recd and
Sent information?
Yes
Go To
Chart 1I
2097937-002 Rev G
11 March 2019
Yes
Verify Communication cable
has not been damaged.
Verify Communication cable
No
is connected at both ends.
T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V
Download the appropriate drive based on
your computer OS from the Flash Drive
supplied with T2100-ST Series Treadmill.
When connecting to the USB port, ensure port
configuration matches with your software
port identification.
After installing USB driver and configured
the port setting, reference 1H.
Replace Communication
Cable.
Communication
should be restored
(Reference Chart 1G).
Connect Communication
Cable.
Communication should
be restored (Reference
Chart 1G).
Troubleshooting
Yes
Is
Communicatio
n Cable
Damaged?
No
No
Is
Communicatio
n Cable
Connected?
Yes