Calibrating Elevation

NOTE:
Tools Required
o
o
o
o
o
o
o
o
1.
2.
3.
4.
5.
2097937-002 Rev G
11 March 2019
Calibrating Elevation is required if a Smart Power Supply or
Elevation Actuator was replaced, or if reported elevation is
not achieving desired accuracy.
Calibration Program (T2100-ST Series Calibration Software)
Tape Measure
¼" Nut Driver or Flat Headed Screw Driver
T15 Torx Screw Driver
Phillips Head Screw Driver
Needle Nose Pliers
Computer or laptop
Serial or USB Communication Cable
Approximate time to complete 30 min
If not already using it, open the T2100-ST Series Calibration
Software.
In the lower left of the window, select the appropriate COM and
Open
press
.
For example:
When connected, the numerical count to the right of Sent: will
increase in value:
This is the indication that the communication has been
established with the treadmill.
Grade Cal
Select the
Test 0.5%
Select
to elevate the treadmill 0.5% and then cycle
the treadmill power.
By resetting the power, the treadmill will automatically find the
lower limit position.
Using a tape measure, verify the 0% elevation.
Refer to the "Elevation Chart" on page 174.
a.
If the zero position requires adjustment, remove actuator
top cover.
VonWeise Actuator
T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V
.
tab.
MSI Actuator
Calibration
171