Troubleshooting
Smart Power Supply Error Code 3 Flow Chart 1L "Zero Speed
Motor Controller Error (3)"
1L
Is the Red LED D13
flashing 3 times with 4
second pause on the
Smart Power Supply
PCBLF0443-1?
Yes
Turn the treadmill power
switch OFF (O).
Verify the harness between Drive PC2303-012-N
and Smart Power Supply PCBLF0443-1 is properly
connected and not damaged.
Verify the harness between Motor
00101-03 and Drive PC2303-012-N is properly
connected and not damaged.
Replace Drive
PC2303-012-N
Return Treadmill
to service.
90
No
PMC5FA-
Send a Start Belt Command or
Press Start if equipped with
controller.
Did the
No
running belt
start?
Yes
T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V
Yes
Go To
Page 1I
Is the Red LED D13
flashing on the Smart
Power Supply
?
PCBLF0443-1
Was the harness
Yes
damaged?
No
Was the harness
No
properly
connected?
Yes
Was the harness
Yes
damaged?
No
Was the harness
No
properly
connected?
Yes
Turn the treadmill
power switch ON (-).
Is the Red LED D13 on
Yes
the Smart Power
Supply PCBLF0443-1
on continuously?
No
Treadmill should
be operational.
Replace Harness
10 Pin Harness
Connect harness.
Replace Motor
Connect harness.
No
Go To
Page 1I
2097937-002 Rev G
11 March 2019