5.
6.
7.
8.
9.
10.
11.
12.
13.
2097937-002 Rev G
11 March 2019
Fig. B
With the flat head screwdriver, remove the ESB and STS
assembly from the fabricated handrail (Fig. B).
Caution: Use ESD precautions including wearing a
grounding wrist strap which is connected to an exposed
metal connection point on the treadmill chassis.
Completely remove the ESB and STS assembly and gently
pull to expose the wires.
With two pairs of long nose pliers, gently separate the
connectors on the ESB.
With a pair of pliers, remove the STS connector from the
tether assembly (Fig. C).
Fig. C
Fig. D
Reverse steps 1 through 8 to reinstall the ESB and STS
assembly.
Remove the (5) hood-securing bolts.
Remove the hood and set it aside.
Swap J4 and J5 with J6 and J7 on the controller board.
With the pliers, cautiously remove the wires. Slowly wiggle
the wires up and down while pulling. Avoid causing
excessive flexing to the board (Fig. D).
Replace the hood and secure it with the original screws.
T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V
Component/FRU Removal and Replacement
J6 & J7
J4 & J5
Screwdriver
Fabricated
Handrail
STS Connection on
outer spade blades.