Mac 2000 St Connection And Setup - GE T2100-ST1 Service Manual

Hide thumbs Also See for T2100-ST1:
Table of Contents

Advertisement

Assembly and Setup
1.
2.
3.
4.
5.
6.

MAC 2000 ST connection and setup:

NOTE:
54
MAC 5500 ST (Keyboard)
Turn the power OFF at the treadmill.
Connect the 9-pin D-Sub end of the communication cable
(GE part number 2097829-108) to the RS232 COM port
located on the lower front panel of the treadmill.
Connect the 8-pin mini DIN of the communication cable (GE
part number 2097829-108) to the COM1 port on the
MAC 5500 ST unit.
MAC 5500 Rear View
Make sure the treadmill is connected to AC power.
Turned the power ON at the treadmill.
If necessary, turn on the MAC 5500 ST device with the
power button on the keyboard.
Refer to the MAC 5500 operator's manual for instructions
on conducting an exercise stress test.
Prior to starting this section, confirm that the MAC 2000
device contains the ST (Stress Test System) option. The
device can be identified by the two additional rows of keys
on the keyboard, as shown (1) in the following illustration.
T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V