GE T2100-ST1 Service Manual page 169

Hide thumbs Also See for T2100-ST1:
Table of Contents

Advertisement

8.
9.
2097937-002 Rev G
11 March 2019
seconds per calibration point, with the overall time of the
speed calibration procedure taking up to 5 minutes to
complete.
When calibration routing is complete the treadmill will stop
and, if successful, the Speed Calibration table will have
automatically updated all of the values in the Measured
Speed column (for example see table below). At this point
running belt calibration is complete.
NOTE:
If the calibration routine fails, refer to the flow chart
titled Smart Power Supply Error Code 1 Flow Chart 1J
"Bad Calibration Error (1)" on page 87.
Select the Basic Control tab, select Start Belt, and then
press
one time.
The running belt will start at 0.1 mph.
NOTE
You have two ways to change speed:
selecting
speed, or
double-clicking on the Set Speed and typing in
the desired speed
To validate the belt speed calibration, do the following:
a.
Verify a point on the run belt that can be seen plainly with
each full revolution; this could be the seam or a chalk line
you make on the belt.
T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V
or
and adjusting to the desired
Calibration
169