Table 2: Guidance and Manufacturer's Declaration –
Immunity All ME Equipment and ME Systems
Immunity
Test
ESD
IEC 61000-4-2
EFT
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage
Dips/Dropout
IEC 61000-4-
11
Power
Frequency
50/60Hz
Magnetic
Field
IEC 61000-4-8
2097937-002 Rev G
11 March 2019
The T2100-ST Series is intended for use in the
electromagnetic environment specified below. The
customer or user of the T2100-ST Series should ensure
that it is used in such an environment.
IEC 60601
Test Level
±8kV Contact
±15kV Air
±2kV Mains
±1kV I/Os
±1kV
Differential
±2kV
Common
>95% Dip for
0.5 Cycle
60% Dip for
5 Cycles
30% Dip for
25 Cycles
>95% Dip for
5 Seconds
30 A/m
T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V
Compliance
Electromagnetic
Level
Environment - Guidance
±8kV Contact
Floors should be wood,
concrete or ceramic tile.
±15kV Air
If floors are synthetic, the
r/h should be at least
30%
±2kV Mains
Mains power quality
should be that of a
±1kV I/Os
typical commercial or
hospital environment.
±1kV
Mains power quality
Differential
should be that of a
typical commercial or
±2kV
hospital environment.
Common
>95% Dip for
Mains power quality
should be that of a
0.5 Cycle
typical commercial or
hospital environment. If
60% Dip for
the user of the T2100-ST
5 Cycles
Series requires continued
operation during power
mains interruptions, it is
30% Dip for
recommended that the
25 Cycles
T2100-ST Series be
powered from an
uninterruptible power
>95% Dip for
supply or battery.
5 Seconds
30 A/m,
Power frequency
50/60Hz
magnetic fields should be
that of a typical
commercial or hospital
environment.
Introduction
15