Honeywell EK280 Applications Manual page 64

Volume converter
Hide thumbs Also See for EK280:
Table of Contents

Advertisement

EK280
a) Close time window / pre-allocate connections
To ensure that parameterization functions correctly, the time windows 1 to 4 need to be closed, and
parameterization of the connections must be pre-allocated. To do this:
Connect a PC or laptop with enSuite installed to the EK280 optical interface using the connection
cable.
Start enSuite and set up the communication link to the EK280 using the optical interface.
( chapter
3.1 Setting up the communication link to the EK280 via optical
Send the following files to the EK280 ( chapter
types of connections and time windows to factory setting (*)
Then wait a minute, for any communications activity in progress on the EK280 to be safely stopped.
b) Sending parameter file
Depending on built-in modem type ( Kapitel 7.1.1) send one the following files to the EK280
( chapter
3.6 Send parameter
- ECM-GW120: EK280_xvx.. - int. ECM-GW120 - WipSoft (GSM) (*)
- ECM-2G-UG350:
- ECM-3G-UU270:
For the following customized settings and tests required, press "Navigation" on the left-hand side
and then
"Parameterize
i
All the values listed in enSuite under "Parameters" can be found on the "Serv." display tab of the
EK280 under the same path.
c) Setting time window for potential data retrievals
In the base configuration now displayed you can define cyclic time windows, during which data can be
retrieved from the EK280 via the modem. There are two available time windows, which operate on EK280
battery power only and for versions of the EK280 with power supply there are two more, which operate only
on mains voltage.
Make the battery power time windows only as long and frequent as absolutely necessary. Each
i
time window reduces the EK280's battery life.
The editable time windows are not directly accounted for in calculation of expected battery life of
the "Bat.R" EK280 batteries. Therefore the expected battery life is not updated, when the time
windows are changed. Past time windows and data transfers are however used in remaining
battery and service life calculation for the "BatRC".
The two battery-power time windows are set in the enSuite device window under:
-
Parameters > Interfaces> Connections > Connection 1
-
Parameters > Interfaces> Connections > Connection 2
4
The "x" stands for the version of software on the device, e.g. "2v2".
The star * stands for a small letter which indicates the version of the file, e.g. "a".
Application Manual
file, p. 21):
EK280_xvx.. - int. ECM-2G-UG350 (GPRS) (*)
EK280_xvx.. - int. ECM-3G-UU270 (GPRS) (*)
online" at the bottom of the navigation window.
3.6 Send parameter
file, p.
4
4
4
4
and
62
interface, p. 19)
21):EK280 - Reset all