Clear Archives ("Clr.a"); Parameterize The Volume At Measurement Conditions "Vm - Honeywell EK280 Applications Manual

Volume converter
Hide thumbs Also See for EK280:
Table of Contents

Advertisement

EK280
Press the ENTER key. ⇒
Change the value using the arrow keys  or  to "1".
Press the ENTER key to confirm the set value. It is possible to abort the entry by pressing the ESC
key.

2.4.3 Clear archives ("Clr.A")

i
With the
"Clr.A"
archive, monthly archives 1 and 2 and the four user archives can be deleted.
So that the measuring period archive, the daily archive, monthly archives 1 and 2 and the four user
archives are deleted from the device, go to and open the administrator or calibration lock as
described in the  EK280 Operating Instructions [1].
Move the cursor in tab
Press the ENTER key. ⇒ One digit of the value flashes.
To prevent accidental deletion of meter counts readings and archives, a security function is provided.
To execute the delete function, the serial number of the device must be entered. For input, move the
cursor with the arrow keys  or  to the digits and change these with the arrow keys  or .
i
The serial number of the device is printed on the identification label or it can be fined in the
"Serv."
register under
Press the ENTER key to confirm the value setting. Input can be canceled using the ESC key.

2.4.4 Parameterize the volume at measurement conditions "Vm"

i
A one-time setting up of the volume counter while the administrator lock is open is possible from
software version 2.20 and higher.
The counter of the EK280 can be set once in a similar way to that of the gas counter while the administrator
lock is open for recording the volume in practice. When the calibration lock is open it is always possible to
set the volume and this can also be done directly via the operational volume (tab
"Actual volume" >
"Vm".
Method (without opening the calibration lock):
Open the administrator lock.
Move the cursor to
to be taken over, is first interim saved by the "resettable interim counter" VmP and then taken over
as "Vm" when the counter reading is reached).
Press the ENTER key. ⇒ One digit of the value flashes.
Move the cursor with the arrow keys  or  to the digits and change these with the arrow keys 
or .
Application Manual
"0"
flashes.
functions described here, only the measurement period archive, the daily
to
"Serv."
"Edit and delete > "Clr.A"
"Identification" > "Volume Converter" >
"Serv."
>
"Volume" > "Actual volume" > "Synchronization Vm > "VmA"
.
"SNo".
"Serv."
to
"Volume" >