Hitachi CHROMASTER 5160 Instruction Manual page 97

Table of Contents

Advertisement

PROG
(SET PROG No.)
1
No. of program containing parameter to
be deleted (m)
(MODIFY=0, NEW=1)
3
(MAX)
Upper limit pressure
level
5
6
ENT
(8)
Deletion of Parameter
The parameter deletion method is shown next.
2
0
ENT
7
DEL
ENT
1
Program number can be from 1 to 9.
2
For correction, input 0. For generating program, input 1.
3
If pump pressure rises beyond the level set here, the pump
automatically stops.
Values that can be inputted are 0.1 to 60.0 MPa.
4
If pump pressure falls below the level set here, the pump
automatically stops.
Values that can be inputted are 0.0 to 60.0 MPa.
5
Press the
the screen. If display advances beyond that step, press
the
key.
6
Move the cursor to the parameter to be deleted while
holding down ENT key.
7
To delete the parameter, press the DEL key and then
press ENT key.
8
When the ESC key is pressed, Edit is finished, and
Monitor screen appears.
ENT
ENT
4
(MIN)
Lower limit pressure
level
8
ESC
key until the step to be corrected appears on
3 - 60
ENT