3.4 Analyzing Operation
(b)
Storing the Background Spectral Data in Memory ("2" key
followed by ENT key)
Scan is automatically carried out.
During acquisition of the background spectrum, the
following screen is presented.
SCANNING BACK GROUND SPECTRUM
WL: *** nm
Upon completion of spectral data acquisition, the monitor
screen automatically appears.
(c)
Storing the Reference Energy with D
key followed by ENT key)
Set the minimum wavelength of scan range.
SCANNING WL RANGE (WL1 WL2) nm
<200 – 850>
Press [numerical] keys (within [2][0][0] to [8][5][0]), and then
the [ENT] key.
* Enterable range: 200 to 850 nm, enterable in 1-nm
steps
Set the maximum wavelength of scan range.
SCANNING WL RANGE (WL1 WL2) nm
<250 – 600>
Press [numerical] keys (within [2][5][0] to [9][0][0]), and then
the [ENT] key.
* Enterable range: Minimum scan range is 50 nm.
Measuring span is max. 400 nm.
Input the number of spectral data.
MEMORIZE SPECTRUM NO. (1–3)
_
3 - 32
lamp in Memory ("3"
2
D2 REF
200 – 600
D2 REF
200 – 600