COM
Lamp igniter
Cathode cable
Grip
Fig. 5-4
(f)
Injury due to Xenon Lamp Explosion (at disposal)
The xenon lamp is filled with high-pressure gas (approx.
1 MPa at room temperature, approx. 4 MPa under
operating condition), and this high-pressure gas still
remains in the xenon lamp after it is demounted for
replacement.
For disposal of the used xenon lamp, wrap it with a thick
cloth (e.g., triple-folded cotton cloth) completely and crush
its glass part with a hammer or the like. Then, discard the
xenon lamp as a dangerous waste item properly.
If the xenon lamp is discarded without being crushed, it
could explode due to possible impact at the time of
disposal, scattering glass pieces to cause personal injury.
Injury due to Xenon Lamp Explosion (at mounting)
When loosening or tightening the retaining nut for the
xenon lamp, be careful not to apply excessive force to its
glass bulb part. Never hold and turn the glass bulb part
for loosening/tightening.
+HV
Lamp House after Removal
Removal of old xenon lamp
WARNING
WARNING
5 - 7
Anode cable
Xenon lamp

Advertisement

Table of Contents
loading

Table of Contents