Appendix 1.3 Items To Be Prepared By Customer; Appendix 1.3.1 Power Supply Arrangement - Hitachi CHROMASTER 5160 Instruction Manual

Table of Contents

Advertisement

Appendix 1.3 Items to be Prepared by Customer

Appendix 1.3.1 Power Supply Arrangement

Confirm that the following requirements are satisfied before
installing the instrument.
(1)
Power Supply
Power supply
voltage
Frequency
Power capacity : 100 VA or more for a single instrument,
Power outlet
APPENDIX 1 - 4
: 100 to 240 V AC
Fluctuation shall be within ±10% of the
rated voltage.
: 50 or 60 Hz
Fluctuation shall be within ±0.5 Hz of the
rated frequency.
and 600 VA or more for the total system
The total system shown comprises
Chromaster organizer, Chromaster 5110
pump, Chromaster 5210 autosampler,
Chromaster 5310 column oven, and
Chromaster 5410/5420/5430/5450
detector.
Check the necessary power supply
requiretemnts and prepare a power
supply meeting the specified
requirements taking into account your
actual system configuration and the OA
equipment to be used in combination.
: Power supply shall be obtainable near
the place for installing the system
inclusive of this instrument.
This detector requires at least one
grounded power outlet (3-pole type).
At least one additional outlet is needed
for the entire system. (See section
2.5.2.)

Advertisement

Table of Contents
loading

Table of Contents