2.4 Specifications

2.4.1 Main Unit

(1)
Maximum number of samples
120 (1.5 mL)
72
(4 mL) (option)
195 (1 mL) (option)
192 (96-holes microplate × 2) (option)
768 (384-holes microplate × 2) (option)
(2)
Sample injection method
Cut injection method
All volume injection method
Full loop injection method
(3)
Syringe capacity
175 µL (standard)
70 µL (option)
(4)
Sample injection volume
0.1 to 100 µL
(5)
Loop capacity
100 µL (standard)
200 µL (accessory)
5, 10, 20 µL (option)
(6)
Syringe speed
5 steps
(7)
Reproducibility of injection volume*
≤ 0.2 % RSD (in 10 μL injection, Cut injection method)
≤ 0.25 % RSD (in 5 μL injection, Cut injection method)
≤ 0.9 % RSD (in 1 μL injection, Cut injection method)
≤ 1.0 % RSD (in 1 μL injection, All volume injection method)
≤ 0.2 % RSD (in 5 μL injection, Full loop injection method)
(8)
Carryover
≤ 0.003 % (Cut injection method)
(9)
Linearity of injection volume
Correlation coefficient: ≥0.999 (Cut injection method)
(10) Injection volume accuracy
± 2 %
2 - 44