3.4 Analysis Operation
3.4.1 Time Program Setting
SET PROGRAM NO. (1 - 9)
1
EDIT MODE (MODIFY=0,NEW=1)
0
TIME
Time setting
0.0
TIME
Temperature
setting
0.0
TIME
Valve
setting
0.0
TIME
New step
_
In the Chromaster 5310, a time program can be prepared for
each of the temperature setting and selector valve position
setting.
PROG
Enter a time program number.
1 ENT
Select a mode.
• MODIFY :
• NEW
0 ENT
TEMP VALVE1
Enter a preset time.
ENT
TEMP VALVE1
_
Enter a preset temperature.
40 ENT
TEMP VALVE1
40
_
2 ENT
TEMP VALVE1
For the 2nd step onward, it is not allowed to enter 0.0 min.
On completion of program settings, press the ESC key.
Monitor screen
For editing an existing time program.
:
For deleting an existing time program and
creating a new one.
1
1
1
EVENT
1
3 - 17
3.4.1
Indicates the number
assigned to the time program
being edited.