3.1.3 Temperature Setting
Carry out setting of the control temperature and temperature
upper limit.
SET TEMP (°C)
<1 - 85>
SET TEMP (°C)
<45 - 95>
NOTE: 1. Unless a control temperature and temperature upper
limit are entered (and ENT key clicked) on the
temperature setting screen, the settings will not be
updated.
2. The temperature upper limiter should normally be set
at 70 °C or higher except for a special reason.
If the limiter were erroneously set near the ambient
temperature, then a temperature limiter error may be
caused by ambient temperature fluctuation.
3. If a temperature upper limiter error occurs, the
temperature control will be stopped automatically.
To resume the temperature control, make resetting
on the temperature setting screen.
3 - 7
SET TMP
MAX LIMIT (°C)
40
70
Set the control temperature.
40 ENT
MAX LIMIT (°C)
40
70
Set the temperature upper limit.
70 ENT