Waste solution bottle
Degassing
Detector (wavelength 265 nm, time constant 0.1 s)
Oven (40 °C)
Autosampler
Pump
unit
Fig. 4-1 Flow Path Diagram
(2)
Connect the e-Line cable among the autosampler, detector
and pump.
(3)
Check if the drain tube is properly connected to the waste
solution bottle. The waste solution bottle must be at a
lower position than the autosampler.
(4)
Prepare other units to be connected to the autosampler as
described in the respective instruction manuals.
(5)
Put the drain tube from the detector into the exclusive
detector waste liquid bottle. If this tube is put into other
waste solution bottle such as for Chromaster 5210
autosampler, baseline fluctuation may be caused due to
unstable back pressure.
4 - 3
Waste solution bottle
Load resistance coil of stainless pipe
0.25 mm ID × 20 m
(Injection valve)
注入バルブ)
(Wash pump)
Mobile
Wash
phase
fluid