2.4 Specifications
2.4 Specifications
(1)
Principle of Temperature Control
Peltier heating/cooling control
(2)
Temperature Setting Range
1 to 85 °C (in increments of 1 °C)
(3)
Temperature Control Range
[Ambient temperature - 15] to [Ambient temperature + 60] °C
and within temperature setting range
(4)
Temperature Setting Accuracy
Oven temperature
Ambient temperature : ±2.0 °C (in the range of 4 to 35 °C)
(5)
Temperature Control Precision
SD ≤ 0.2 °C (measured at the specified point in the oven)
(6)
Safety
(a) Overheat preventing mechanism by means of a
thermal protector (105 °C)
(b) Organic solvent vapor leak sensor
Temperature control is forced to stop upon detection of
leakage of organic solvent vapor. Leak detection
level setting is allowed.
(c) Door sensor
Temperature control is forced to stop when the door is
opened.
(7)
Power Requirements/Power Consumption
100 V (50 Hz/60 Hz), 230 VA
(when the selector valve (option) is mounted)
2 - 10
: ±1.0 °C (20 to 85 °C in preheating)