4.2 Confirmation of Performance Specifications
(1)
After turning on power supply, set wavelength at 250 nm
and conduct warm-up for about 1 hour.
WL
Press
key.
WAVELENGTH (190 – 600) nm
250
Press [numerical] keys [2][5][0], and then the [ENT] key.
(2)
Carry out auto zero.
A/Z
Press
key.
TIME
WL
0.0
250
(3)
Start the chromato-integrator and record the baseline for 15
minutes.
(4)
From the recorded chart, figure out noise value according to
ASTM1657-98 (JAIMAS 0005-2008).
R=n
Noise value =
R=l
(Excerpt from JAIMAS 0005-2008)