(iii) Press the WL key and the wavelength setting
screen appears.
Set an excitation (Ex) wavelength.
<Wavelength setting screen (Ex indicated)>
Ex
Em
550
350
Press numeric keys 5 5 0 and then the ENT key.
(iv) Emission (Em) wavelength will not be set here.
Leave the present setting as it is.
<Wavelength setting screen (Em indicated)>
Ex
Em
(250 - 900.0)
550
350
Press the ENT key.
(v)
The 1
st
<1
monitor screen>
TIME
Ex
0.0
550
(vi) In this status, place a paper strip at the Ex beam
exit center of the sample compartment as shown
in Fig. 5-9 and observe the condition of the
excitation beam.
Repeat the VERTICAL, HORIZONTAL and
FOCUS adjustments in this order little by little
until the excitation beam (rainbow color) becomes
brightest. The coarse adjustment of the light
source position is thus completed. Please adjust
VERTICAL first in installation because the lamp
position is shifted downward for fixation of the
transport screw.
5 - 15
(200 - 850.0)
st
monitor screen appears.
Em
FL
PROG
350
****
1
LAMP
[I]
Xe