Hitachi CHROMASTER 5160 Instruction Manual page 838

Table of Contents

Advertisement

Syringe vlave
4
Plunger
wash
va/2 + vw/2
+ vi + vd
Syringe
Wash
fluid1
(7)
Buffer tube
Injection valve
3
2
1
5
Wash pump
P
3-way solenoid valve
Degassing
unit
vd :Dead volume (the instrument inside parameter)
vi :Injection volume (the sample net volume to the flow path)
vw :Waste volume (the air volume of the lead and rear part of a sample)
va :Air volume (the air volume which sample is sandwiched.)
Wash
fluid2
Sample loading
Deliver the sample into the sample loop.
The syringe discharges [va/2 + vw/2 + vi + vd : one-half of the
Air volume + one-half of the Waste volume + Injection volume
+ Dead volume (inside parameter)], and delivers a sample
into the loop.
vw/2
5
Loop
6
Waste
1
solution
From Pump To Column
2 - 42
Needle
vw/2
Injection port
4
3
Waste
2
solution
Wash port
Rack
Sample