Hitachi CHROMASTER 5160 Instruction Manual page 967

Table of Contents

Advertisement

(4)
Fill a beaker with wash fluid (e.g. 2-propanol), and put the
solvent filter in it. Perform ultrasonic cleaning for approx. 5
to 6 minutes.
Beaker
(5)
Then, replace the wash fluid with distilled water, and
perform additional ultrasonic cleaning for approx. 5 to
6 minutes.
(6)
After removing water completely, attach the solvent filter to
the inlet tube.
Inlet tube
5 - 24
Solvent filter
Wash fluid
Distilled water
Ultrasonic cleaner
Insert
Solvent filter

Advertisement