3.2.9 Link Program Function

Monitor screen
UTLTY
(SET UP)
UTILITY(CONDITION=1,UNIT=2,FLUSH=3
5
EVENT=4,LINK=5,OTHERS=6)
Select LINK (LINK PROGRAMA)
5
ENT
LINK PROGRAM(YES=1, NO=0)
1
Select YES
1
ENT
LINK RESTART(YES=1, NO=0)
0
1
ENT
0
ENT
Plural programs can be linked.
NOTE:
When connecting with CDS, setting should be set to
NO. In this case, setting will be made on the system
manager side.
(1)
Setting of Link Program
Set the conditions for link program. Explained below is an
example where you set one run of program No. 1 at the
first step, conditioning time at 30 minutes, 3 runs of
program No. 2 at the second step and conditioning time at
20 minutes.
If manual setting screen is displayed, press [INIT] to obtain the
programming mode.
Monitor screen
After executing a link program by a few steps, execution of the next steps will be
avoided and the link program will return to the first step and will be executed
again. (Assume that a 5-step link program is set. If you set link restart after
the 2nd step in this case, the 3rd step will not be executed and execution will
restart from the 1st step.)
Also, when plural cycles of the same step are set, the link program will return to
the first step and will be executed so far as the link restart is set, though all the
set cycles are not completed. Setting should be made when a pump is set in
the initial status ([IN] status is indicated at the top right of LCD). Setting is
impossible during gradient elution and conditioning.