3.3 GLP-Related Functions

Logbooks of D
, W, and Hg lamps can be displayed and reset.
2
And, keys on the keyboard can be locked so that key input will
not be attempted in error during analysis, etc.
GLP
Pressing the GLP key presents the GLP setup screen.
GLP (LOGBOOK=1, KEY LOCK=2,
1
D2 CHECK=3, W CHECK=4, Hg CHECK=5)
Input Numeral
1
LOGBOOK :
2
KEY LOCK :
3
D2 CHECK :
4
W CHECK :
5
Hg CHECK :
3 - 21
(Good Laboratory Practice)
Function
Displays and resets the logbook of
each lamp..
Disables acceptance of keyboard
operation.
Checks energy of the D
250 nm, and wavelength accuracy
using the bright lines at 486 and 656
nm of this lamp.
Checks energy of the W lamp at
600 nm.
Checks wavelength accuracy using
the bright lines at 254 nm of the
mercury lamp.