Hitachi CHROMASTER 5160 Instruction Manual page 560

Table of Contents

Advertisement

Message
LAMP OFF BY
EXTERNAL ERROR
PRESS "CL" KEY TO
TURN ON.
LAMP OFF BY
LEAKAGE
PRESS "CL" KEY TO
CLEAR MESSAGE.
Xe LAMP ERROR
PRESS "CL" KEY TO
CLEAR MESSAGE.
Hg LAMP ERROR
PRESS "CL" KEY TO
CLEAR MESSAGE.
XE LAMP EXCEEDED
EXCHANGE
STANDARD TIME
PRESS "CL" KEY TO
CLEAR MESSAGA
A/Z OVER RANGE
PRESS "CL" KEY TO
CLEAR MESSAGE.
NO SPECTRA
PRESS "CL" KEY TO
CLEAR MESSAGE.
SPECTRA ARE
MISMATCHED
PRESS "CL" KEY TO
CLEAR MESSAGE.
DATA BUFFER
OVERFLOW
PRESS "CL" KEY TO
CLEAR MESSAGE.
THERMO CELL IS NOT
CONNECTED
Description
Lamp has been turned off due
to an external error. Busy
status is assumed.
Liquid leakage has occurred.
Xe and Hg lamps have been
turned off. Busy status is
assumed.
Xe lamp fails to come on or go
off. Temperature upper limiter
of Xe lamp power source is
activated.
Hg lamp fails to come on or go
off.
The cumulative turn-on time of
xenon lamp has exceeded its
standard exchange time.
Allowable auto zero range has
been exceeded.
Spectrum of specified No. is not
found.
Scan mode incorrect,
Em bandpass incorrect, fixed
wavelength incorrect,
background WL range doesn't
include entire spectrum WL
range, background spectrum
mode incorrect, or energy
spectrum incorrect.
During data acquisition, data
transfer to the interface board
has been disabled, causing an
overflow of the data buffer.
A thermo cell is not connected
even though the thermo cell
setting is in ON state.
4 - 34
Remedy
1) Release the key lock if it has
been activated.
2) Press CL key and light up the
lamp again.
1) Press CL key, and 1st monitor
screen returns.
2) Use UTLTY key to light up the
lamp again.
1) Press CL key, and 1st monitor
screen returns.
2) Wait at least 5 minutes after
turning off lamp for the lamp to
cool down, then turn on power
again or re-light the lamp.
3) Replace lamp with a new one.
1) Press CL key, and 1st monitor
screen returns.
2) Replace the lamp with a new
one.
1) Press CL key, and 1st monitor
screen returns.
2) Replace the lamp with a new
one.
1) Press CL key, and 1st monitor
screen returns.
2) Change the wavelength.
3) Replace the solvent with another
one.
4) Clean the flow cell.
1) Press CL key, and 1st monitor
screen returns.
2) Enter a number corresponding to
an existing spectrum.
3) Perform spectrum acquisition.
1) Press CL key, and 1
screen returns.
2) Select correct parameters.
1) Press CL key, and 1
screen returns.
2) Check the connection with the
interface board.
1) Press CL key, and 1
screen returns.
2) Check the thermo cell
connection.
st
monitor
st
monitor
st
monitor