Pressure Monitor - Hitachi CHROMASTER 5160 Instruction Manual

Table of Contents

Advertisement

3.3.6 Pressure Monitor

The total number of times of the double speed section error is
indicated.
The double-speed section error indicates that the pressure fails
to return to the reference level (previous pressure) even after
double-speed control (an increase of pump motor speed) was
effected to compensate for decreased pressure. Air bubbles in
the flow path may cause the error.
Monitor screen
GLP
GLP
(LOGBOOK=1,
2
Select PRESSURE M
2
ENT
PRESSURE MONITOR
9999 TIMES
The total number of times of errors
ENT
or
Monitor screen
The pressure monitor will be reset in the following cases.
▪ Input of PUMP ON or PUMP OFF
▪ Activation of contact
▪ Start of gradient elution
▪ Receiving of start signal from integrator in e-line configuration
▪ External start of pump
▪ Input of PURGE ON or PURGE OFF
3 - 46
PRESSURE M=2,
ESC
KEY LOCK=3)