Advertisement

3. Items to be Prepared by Customer
3.1 Power Supply Arrangement
(1)
Power supply voltage: 100 to 115 V AC/220 to 240 V AC
Fluctuation shall be ±10% of the rated voltage.
(2)
Frequency: 50 or 60 Hz
Fluctuation shall be ±0.5 Hz of the rated frequency.
(3)
Power capacity:
The above total system is an example system configured of
Chromaster organizer, Chromaster 5110 pump, Chromaster
5210 autosampler, Chromaster 5310 column oven and
Chromaster 5410/5420/5430/5450 detector. Confirm the
necessary power capacity taking actual system
configuration and OA equipment in the installation room into
account and prepare a power supply having an adequate
power capacity.
(4)
Power outlet:
Power supply shall be obtainable near the place for
installing the system inclusive of this detector.
This detector requires a single grounded power outlet (3-
pole type). One more outlet is needed for the entire
system.
(5)
Grounding terminal:
Shall have a grounding construction for a ground resistance
of 100 Ω or less.
(a) Usual connection method (with grounded 3P power
outlet)
APPENDIX 1 - 3
100 VA or more for this detector alone,
600 VA or more for the total system