Hitachi CHROMASTER 5160 Instruction Manual page 375

Table of Contents

Advertisement

5.2 Lamp Replacement
5.2.2 Replacement of W Lamp
(e) Key in the date of replacement in the format of month-
day-year. For instance, if the lamp is replaced on
March 1, 2010, key in the date as shown below.
[3] → [ENT] → [1] → [ENT] → [2][0][1][0] → [ENT]
Thus, the date of lamp replacement is stored in
memory. At the same time, the cumulative turn-on
time of the D
on count is reset to 1.
NOTE: The logbook data is reset when the numeric value
of the year and [ENT] are keyed in. Therefore,
even if the same year as that of replacement is
indicated, the year must be numerically input
followed by pressing the [ENT] key.
(1)
Part and Tool to be Prepared
• New W lamp
• Phillips screwdriver
(2)
Check the following.
• The power switch of this instrument is turned off.
• The power cord is unplugged from the instrument.
(3)
How to Replace W Lamp
(a) Detach the lamp cover (by removing 2 fastening
screws).
5 - 8
lamp is cleared to zero (0), and the turn-
2
Fastening screw
Lamp cover

Advertisement