3.1 Replacement Procedure

3.1.1 Fuse Replacement

3. REPLACEMENT PARTS

Electric Shock due to Contact with Inside of Instrument
When replacing the fuse with a new one, be careful not to
receive an electric shock. To prevent this, be sure to turn
off the power switch before proceeding to fuse
replacement.
(1)
Item to be Prepared
New fuse (Part No. J821337)
Be sure to use a time-lag fuse having the specified rating
(T6.3AL for 250 V).
(2)
Pre-check Condition
Make sure that the power switch of the instrument is turned
off and that the power cord is disconnected from the
instrument.
(3)
How to Replace the Fuse
(a) Remove the fuse holder by hooking a tool having a thin
and flat tip (such as blade-edge screwdriver) to the
rear side of the fuse holder as shown below.
Fig. 3-1 Removal of Fuse Holder
3 - 1
WARNING
Fuse holder
3.1.1