Appendix 1.5.5 System Tubing
Central hole
Through this hole, the tube from the
pump to the autosampler and that from
the column to the detector are passed.
In the Chromaster series, the inlet tube up to the pump and drain
tube are fastened to the left side face of each unit with tube
clamps. And the tubes are connected from the pump to the
autosampler, column and detector through the central hole of
each module unit.
Drain tray
Hole on the right hand
Through this hole, the tube
from the autosampler to the
column is passed.
NOTE: Please ensure that the detergent vial is located
higher than that the autosampler main unit. If it
is located lower than the autosampler main unit, it
is possible that the instrument is prevented from
achieving sufficient performance due to the
bubble entering the syringe because of the
negative pressure in the tubing.
NOTE: Please ensure that the detergent is connected to
the degassing unit.

APPENDIX 1 - 20