Troubleshooting: Monitors
In this section
•
•
•
•
UV monitor U9-M and UV
detector unit
Problem
The UV module is not
found by the instru-
ment
ÄKTA avant User Manual 29035184 AE
UV monitor and UV detector
Conductivity monitor
pH monitor and pH valve
Pressure monitors
Possible cause and action
Communication problem
•
Contact Service.
The cable between the UV module and the ICU is
•
not connected
Remove the UV module and make sure that the cable
is connected.
Wrong Node ID
•
Check the module´s Node ID. If necessary, change the
Node ID. See Section 9.15 Check and change the Node
ID of a module, on page 515.