4 Optional instrument modules
4.7 Loop valve
Connect tubing
The table below shows recommended tubing and connectors. For an illustration of tubing
labels see Tubing labels, on page 432.
Tub-
Connection
ing
label
L1
Injection valve posi-
tion LoopF to
Loop valve position
F
L2
Injection valve posi-
tion LoopE to
Loop valve position
E
Connect a Loop valve
The Loop valve is connected to the Injection valve instead of a loop, as described below.
Step
1
2
3
Note:
126
Tubing
ÄKTA avant
ÄKTA avant
25
150
PEEK,
PEEK,
o.d. 1/16",
o.d. 1/16",
i.d. 0.50 mm
i.d. 1.0 mm
PEEK,
PEEK,
o.d. 1/16",
o.d. 1/16",
i.d. 0.50 mm
i.d. 1.0 mm
Action
Connect port E on the Loop valve to port LoopE on the Injection valve.
Connect port F on the Loop valve to port LoopF on the Injection valve.
Connect one or many loops to the Loop valve. See Section 5.7 Sample appli-
cation, on page 189.
Note:
Always use the the first positions of the valve for the connected loops (e.g., if
three loops will be used, use port 1F-3F and the corresponding ports 1E-3E)
to avoid cross-contamination.
It is possible to place the Loop valve in other positions in the flow path than
the one described above. However, the volume used for washes will then be
incorrect, just as the system configuration shown in the process picture.
Connector
Fingertight connector,
1/16"
Fingertight connector,
1/16"
ÄKTA avant User Manual 29035184 AE
Tubing
length
(mm)
160
160

Advertisement

Table of Contents