Flow paths in ÄKTA avant
Second Sample inlet valve
The illustration below shows an optional flow path when a second Sample inlet valve is
installed. The flow is directed from the Out port of the second Sample inlet valve (V9-S2
or V9H-S2) to the buffer port BUFF of the Sample inlet valve (V9-IS or V9H-IS) and leaves
the valve through the Out port to the Sample pump. In the software, 14 sample inlets
can now be selected.
Second Inlet valve
The illustration below shows an optional flow path when a second Inlet valve is installed,
connected to Inlet valve A. The flow is directed from the Out port of the Second Inlet
valve (V9-A2 or V9H-A2) to port A7 of the Inlet valve (V9-IA or V9H-IA) and leaves the
valve through the Out port to the System pump A. In the software, 14 A inlets can now
be selected.
ÄKTA avant User Manual 29035184 AE
V9-IS
S2
S3
S1
S7
S6
Out
V9-A2
8
9
7
10
13
12
14
11
Out
4 Optional instrument modules
V9-S2
S9
S4
S8
S14
S5
BUFF
V9-IA
A3
A2
A4
A6
A7
A5
Out
4.3 Extra inlet valves
S10
S11
S13
S12
Out