4 Optional instrument modules
4.6 External air sensors
L9-1.2
Connection
between...
V9-Inj or V9H-Inj
and L9-1.2
L9-1.2 and V9-C or
V9H-C
1
System properties
Follow the instruction to update the system properties.
Step
1
2
3
4
5
122
Cut the spare piece of tubing delivered with the system to correct length.
Action
Open the system properties Edit dialog box.
In the Component types list, click Monitors and sensors.
Select the External air sensor (L9) check box in the Component selection
list.
In the Property list, click Before sample inlet or After injection valve depend-
ing on where the air sensor is placed.
Note:
The available properties are the same, regardless of which type of air sensor
that is used, L9-1.2 or L9-1.5.
Click OK.
Tubing
Tubing
label
1
5L1
For ÄKTA avant
25: PEEK, o.d.
1/16", i.d. 0.50 mm
For ÄKTA avant
150 : PEEK, o.d.
1/16", i.d. 1 mm
1
5L2
For ÄKTA avant
25: PEEK, o.d.
1/16", i.d. 0.50 mm
For ÄKTA avant
150: PEEK, o.d.
1/16", i.d. 1 mm
Connector
Tubing
length
(mm)
For ÄKTA avant
100
25: Fingertight
connector, 1/16"
For ÄKTA avant
150: Fingertight
connector, 1/16"
For ÄKTA avant
100
25: Fingertight
connector, 1/16"
For ÄKTA avant
150: Fingertight
connector, 1/16"
ÄKTA avant User Manual 29035184 AE