9 Reference information
9.2 Module specifications
Item
Resolution
Linearity
Drift
Noise
Operating pressure
Lamp operating time
Flow cells: U9-M
Flow cells: U9-L
428
Description
0.001 mAU
U9-M: within ± 2% at 0 to 2 AU
U9-L: within ± 5% at 0 to 2 AU
U9-M (2 mm cell at 280 nm): ≤ | 0.2 mAU | AU/h
U9-L (2 mm cell): ≤ | 0.2 mAU | AU/h
U9-M: < 0.08 mAU
U9-L: < 0.1 mAU
0 to 2 MPa
U9-M: > 5000 h
U9-L: > 10000 h
Standard:
Optical path length 2 mm
Cell volume 2 µl
Total volume: 11 µl
Option:
Optical path length 10 mm
Cell volume 8 µl
Total volume 12 µl
Optical path length 0.5 mm
Cell volume 1 µl
Total volume 10 µl
Standard:
Optical path length 2 mm
Cell volume 2 µl
Total volume: 30 µl
Option:
Optical path length 5 mm
Cell volume 6 µl
Total volume 20 µl